<h3>Sorensen DCS55-55 M62 DC Power Supply</h3><p>The Sorensen DCS Series power supplies are general purpose power supplies designed specifically for laboratory test and systems applications requiring variable DC sources with good ripple and regulation characteristics. The power supplies are constant…
Description
<h3>Sorensen DCS55-55 M62 DC Power Supply</h3><p>The Sorensen DCS Series power supplies are general purpose power supplies designed specifically for laboratory test and systems applications requiring variable DC sources with good ripple and regulation characteristics. The power supplies are constant current/constant voltage supplies with an automatic crossover feature. The DCS Series models provides up to 3000 watts.</p><p>Input: 208-230 VAC, 47-63 Hz, 3 phase</p><p>Output: 0-55V, 0-55A</p><p>Options: M62-  Adds extended input plug, 220 volt 3 phase input</p><p>
ADDITIONAL SPECIFICATIONS:</p><p> </p><p> 
Input Power (Standard) Voltage: </p><p>208-230 VAC
Frequency:</p><p>47 to 63 Hz
Phases:
Three, 3-wire plus ground
Regulation (Line or Load)
Voltage:0.1% of max. output  voltage
Current: 0.1% of max. output current
Transient Response:
A 30% step load will recover to within 2% of original value within 10ms.
Stability:
+0.05% of set point 8 hrs. after warm-upand at a fixed line, load and temp.
Total Harmonics: 2.5% max.
Resistive:
Constant Voltage (0-100%) -0 - 5k ohms
Constant Current (0-100%)- 0 - 5k ohms
Voltage:
Constant Voltage (0-100%)- 0 - 5/10 VDC
Constant Current (0-100%)- 0 - 5/10 VDC
Current:
Constant Voltage (0-100%)-0 - 1 mA
Constant Current (0-100%)-0 - 1 mA
Remote Sensing: Terminals are provided to sense output voltage at point of load.
</p><p>ENVIRONMENTAL CHARACTERISTICS:</p><p>Temperature Coefficient:
-0.02%/°C of max. output voltage rating for
voltage set point.
-0.03%/°C of max. output current rating for
current set point.
Ambient Temperature:
Operating: 0 to 50°C
Storage: -40° to 75°C</p><p>Cooling: Internal fans
 </p><p>PHYSICAL CHARACTERISTICS:
Width   19in
Depth   18 in
Height  3.5 in
Weight  45lbs max
 </p>
ADDITIONAL SPECIFICATIONS:</p><p> </p><p> 
Input Power (Standard) Voltage: </p><p>208-230 VAC
Frequency:</p><p>47 to 63 Hz
Phases:
Three, 3-wire plus ground
Regulation (Line or Load)
Voltage:0.1% of max. output  voltage
Current: 0.1% of max. output current
Transient Response:
A 30% step load will recover to within 2% of original value within 10ms.
Stability:
+0.05% of set point 8 hrs. after warm-upand at a fixed line, load and temp.
Total Harmonics: 2.5% max.
Resistive:
Constant Voltage (0-100%) -0 - 5k ohms
Constant Current (0-100%)- 0 - 5k ohms
Voltage:
Constant Voltage (0-100%)- 0 - 5/10 VDC
Constant Current (0-100%)- 0 - 5/10 VDC
Current:
Constant Voltage (0-100%)-0 - 1 mA
Constant Current (0-100%)-0 - 1 mA
Remote Sensing: Terminals are provided to sense output voltage at point of load.
</p><p>ENVIRONMENTAL CHARACTERISTICS:</p><p>Temperature Coefficient:
-0.02%/°C of max. output voltage rating for
voltage set point.
-0.03%/°C of max. output current rating for
current set point.
Ambient Temperature:
Operating: 0 to 50°C
Storage: -40° to 75°C</p><p>Cooling: Internal fans
 </p><p>PHYSICAL CHARACTERISTICS:
Width   19in
Depth   18 in
Height  3.5 in
Weight  45lbs max